Technical note: False low turbidity readings from optical probes during high suspended-sediment concentrations

Nicholas Voichick, David J. Topping, and Ronald E. Griffiths

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
ED: Publish subject to minor revisions (review by editor) (08 Jan 2018) by Bettina Schaefli
AR by Nicholas Voichick on behalf of the Authors (09 Jan 2018)
ED: Publish as is (24 Jan 2018) by Bettina Schaefli
AR by Nicholas Voichick on behalf of the Authors (02 Feb 2018)
Download
Short summary
This paper describes instances in the Grand Canyon study area and a laboratory experiment in which very high suspended-sediment concentrations result in incorrectly low turbidity recorded with a commonly used field instrument. If associated with the monitoring of a construction or dredging project, false low turbidity could result in regulators being unaware of environmental damage caused by the actually much higher turbidity.