Articles | Volume 23, issue 6
https://doi.org/10.5194/hess-23-2637-2019
https://doi.org/10.5194/hess-23-2637-2019
Technical note
 | Highlight paper
 | 
19 Jun 2019
Technical note | Highlight paper |  | 19 Jun 2019

Technical note: A microcontroller-based automatic rain sampler for stable isotope studies

Nils Michelsen, Gerrit Laube, Jan Friesen, Stephan M. Weise, Ali Bakhit Ali Bait Said, and Thomas Müller
Abstract

Automatic samplers represent a convenient way to gather rain samples for isotope (δ18O and δ2H) and water quality analyses. Yet, most commercial collectors are expensive and do not reduce post-sampling evaporation and the associated isotope fractionation sufficiently. Thus, we have developed a microcontroller-based automatic rain sampler for timer-actuated collection of integral rain samples. Sampling periods are freely selectable (minutes to weeks), and the device is low-cost, simple, robust, and customizable. Moreover, a combination of design features reliably minimizes evaporation from the collection bottles. Evaporative losses were assessed by placing the pre-filled sampler in a laboratory oven with which a diurnal temperature regime (21–31 C) was simulated for 26 weeks. At the end of the test, all bottles had lost less than 1 % of the original water amount, and all isotope shifts were within the analytical precision.

These results show that even multi-week field deployments of the device would result in rather small evaporative mass losses and isotope shifts. Hence, we deem our sampler a useful addition to devices that are currently commercially available and/or described in the scientific literature. To enable reproduction, all relevant details on hard- and software are openly accessible.

1 Introduction

The stable isotopes 18O and 2H represent nearly ideal tracers that are frequently applied in hydrology and other disciplines. Yet, most applications require data on the isotope composition of precipitation (Bowen and Revenaugh, 2003; Hughes and Crawford, 2013).

Although isotope analyzers have become field-deployable, enabling high-resolution on-site rain analyses (Berman et al., 2009; Herbstritt et al, 2018a; Munksgaard et al., 2011, 2012; Pangle et al., 2013; von Freyberg et al., 2017; Windhorst et al., 2017), many researchers hesitate to take their analyzer to the field due to the risk of damaging the expensive device and due to logistical constraints such as high power demand (Herbstritt et al., 2018b). Hence, most samples are still obtained in more traditional ways for subsequent analysis in the laboratory. They are collected manually (e.g., in intra-event studies; Conroy et al., 2016; Michelsen et al., 2015), with cumulative rain collectors (summarized in Michelsen et al., 2018), or with automatic samplers.

Such automatic samplers, initially mostly developed for precipitation chemistry studies, cover a broad range of designs (see reviews by Krupa, 2002; Laquer, 1990; Robertson et al., 1980). Simple versions of the “linked collection vessels” type (Robertson et al., 1980), featuring a set of serially connected bottles that are consecutively filled, are inexpensive and easy to construct. They have no moving parts and collect samples on a volume basis (Laquer, 1990). Particularly the designs by Kennedy et al. (1979) and Vermette and Drake (1987) are still used nowadays (Fischer et al., 2016, 2017; Hervé-Fernández et al., 2016; Muñoz-Villers and McDonnell, 2013; Saffarpour et al., 2016), albeit sometimes with modifications (Buda and DeWalle, 2009; Qu et al., 2017). Drawbacks of this sampler type comprise the lack of an efficient evaporation reduction mechanism, possibly causing post-sampling fractionation (Fischer et al., 2017), and a potential for cross-contamination of samples.

More complex “automatically segmenting samplers” (Robertson et al., 1980) are commercially available but usually expensive (several thousand euros; Tauro et al., 2018). Further, they often do not minimize evaporation sufficiently (Hartmann et al., 2018; Tauro et al., 2018; Williams et al., 2018). To overcome the latter problem, researchers have occasionally added paraffin oil to the collection bottles of such commercial samplers (Birkel et al., 2011; van Huijgevoort et al., 2016; Sprenger et al., 2017; Tunaley et al., 2017), as recommended by Fergusson (1921). However, a quantitative oil removal is not easy, and oil traces might cause problems during laboratory analyses (Gröning et al., 2012), particularly in laser spectroscopy (IAEA, 2014). Fischer et al. (2016) thus explicitly avoided the oil method and preferred to empty their automatic samplers directly after rainfall events.

The isotope community also developed various custom-made automatic samplers, mostly for high-resolution sampling of rainwater and/or surface water (e.g., Camacho Suarez et al., 2015; Conroy et al., 2016; Coplen et al., 2008; Muller et al., 2015; Neuhaus, 2016, 2018; Siebert, 2015; Terzer-Wassmuth et al., 2018; Zannoni et al., 2019). While some of these collectors are outlined in the corresponding patent specifications (Coplen, 2010; Neuhaus, 2016; Siebert, 2015), several others are only briefly described or not at all. We can imagine that, at least in some of these cases, the researchers did not provide details because they saw their collectors as means to an end, i.e., as tools helping to generate data on which they eventually concentrated.

Notable exceptions in this context are Hartmann et al. (2018) and Nelke et al. (2018), who do provide sufficient details on their devices to enable reproduction. The former authors use a peristaltic pump to inject water directly into airtight vials. Nelke et al. (2018), by contrast, utilize two peristaltic pumps to direct water into aluminum-lined bags using solenoid valves. Both collectors apparently focus on continuously flowing media (e.g., dripwater, surface water; instead of discontinuous media such as precipitation) and take discrete “snapshot” samples (instead of integral samples) of small to moderate size (12 and 250 mL, respectively). Moreover, their designs have in common that they are rather sophisticated. Yet, complexity can be a double-edged sword. While their technical solutions are certainly elegant, the advanced designs might be somewhat difficult to reproduce.

Another noteworthy open-source device is the autonomous rainfall sampler by Ankor et al. (2019), which collects daily and monthly samples. Here, the water flows by gravity through the partly 3-D printed system and is elegantly guided by a split tipping bucket and a “water switch” into the daily and monthly bottles with volumes of 225 mL and 2 L, respectively. Unfortunately, a 13-week field test (mean temperature: 25.3 C) revealed evaporative losses from the pre-filled bottles, resulting in remarkable δ18O changes. The latter ranged from about 1.10 ‰ to 2.63 ‰, depending on the bottle cap material. Only when paraffin oil was used in the collection bottles (see above) were the isotopic changes acceptable (approx. 0.14 ‰ for δ18O). To mitigate the evaporation effect, the authors suggest a coupled hydrologic–isotopic model that takes advantage of the combined daily and monthly sample collection and allows a back-calculation of the original sample volumes and isotopic compositions. Although such an approach might be an option, we think it would be more straightforward to minimize evaporation from the collection bottles in the first place.

Here, we describe a complementary automatic rain sampler. Our simple, robust, and low-cost collector allows the timer-actuated collection of integral rain samples, with time intervals ranging from minutes to weeks. A combination of design features effectively reduces post-sampling evaporation. In the spirit of open science (see Ankor et al., 2019; Hartmann et al., 2018; Nelke et al., 2018), we provide a detailed description of our customizable collector and its components. Moreover, we present the results of a 26-week test addressing the evaporation reduction capacity of the device in a warm climate.

2 Design

The following section gives an overview of the sampler design. Further details (bill of materials, technical drawings, circuit diagram, code, and manuals) enabling reproduction are provided in the Supplement and on the website at https://www.ufz.de/index.php?en=44048 (last access: 15 June 2019) (section Documentation).

The sampler collects rain by a funnel from which the water flows into a distribution unit, the core of the device (Fig. 1). It consists of two custom-made uniaxial discs (separated by a 2 mm neoprene rubber seal) with drill holes that are positioned opposite each other. The upper disc (PTFE) is the rotor (rotates clockwise) and has two drill holes fitted with push-in ports. The outer port is the rain inlet, and the inner port is the air outlet. The lower disc (PVC-U) is the stator (remains static) and features 36 ports that are arranged in two circles. The ports of the outer circle are connected to water tubes, and the ports of the inner circle are connected to air tubes (LDPE; see section “Pre-test of tubing materials” in the Supplement, incl. Fig. S1, Table S1).

https://www.hydrol-earth-syst-sci.net/23/2637/2019/hess-23-2637-2019-f01

Figure 1Overview of the automatic sampler principle (photo by André Künzelmann, UFZ). Water (blue arrows) flows from the rain inlet through the distribution unit and into the first bottle (cutaway view). Displaced air (red arrows) leaves the bottle and flows through the distribution unit and the air outlet into a 15 m long pressure equilibration tube. At the end of the sampling interval, the rotor is turned (grey arrow), leading the rain into the second bottle.

Download

Water coming from the funnel flows through the rain inlet into the distribution unit and then through a water tube, into the first 500 mL sampling bottle. All tubes are guided through the bottle caps by means of cable grommets (Fig. S2), ensuring a tight connection. With respect to the bottles, we recommend thick-walled HDPE bottles that effectively reduce diffusive water losses (Spangenberg, 2012; Gröning, 2019). For our purposes, we selected 500 mL HDPE wide-mouth bottles by Labsolute (Renningen, Germany; wall thickness approx. 1.7 mm).

To minimize post-sampling evaporation, we adapted the concept of an evaporation-free cumulative collector (Gröning et al., 2012), which has been successfully tested under hot and arid conditions (Michelsen et al., 2018) and is advocated by the IAEA (2014). In this concept, an inlet tube extends to the bottom of the bottle. A few millimeters of rainfall is sufficient to cause a water level rise into the tube, thus decoupling the bottle headspace from the atmosphere. This process can be amplified by inserting the end of the inlet tube into a small container (Gröning et al., 2012; or a short piece of bent tubing; see Fig. 1), reducing the amount of rain needed to decouple the air in the bottle from the atmosphere. The air displaced by the inflowing water leaves the bottle through the air tubing that, unlike the inlet tube, extends just below the bottle cap. The air tubing leads from the bottle cap back to the distribution unit. Here, the air is pushed through the air outlet into a 15 m long pressure equilibration tube (see Gröning et al., 2012; IAEA, 2014).

At the end of the freely selectable sampling interval, a microcontroller (Texas Instruments; for details, see Supplement) triggers a stepper motor to turn the rotor of the distribution unit by 20. Through this rotation, the first sampling bottle is isolated from the atmosphere – the rotor disc closes the water and air tubes of this bottle. In turn, rain can now flow into the second bottle. The air thus displaced from the second bottle is pushed into the aforementioned pressure equilibration tube; i.e., this tube is shared by all bottles. A tight fit of the rotor is ensured through a spring above the disc, providing sufficient pressure to avoid leakages.

The microcontroller features a display and buttons for convenient programming, avoiding the necessity of a notebook during field setup or maintenance. Moreover, the microcontroller has a set of low power modes that are excessively used. In fact, the microcontroller and the accompanying stepper motor driver are only active during initial setup and between sampling periods, while the distribution unit is driven to the next sampling port. The fact that the distribution unit consumes virtually no power during sampling allows for week-long sampling periods on just two AA batteries for the logic unit and another eight AA batteries for the stepper motor.

The stepper motor and the control unit (microcontroller with accessories and batteries) are each located in a dust- and waterproof enclosure and connected by a waterproof cable (IP68). This robust design loosens the requirements for the overall enclosure. For field deployments (not presented here), a simple plastic storage box was chosen to be sufficient to house the sampling bottles and the automatic collector.

3 Evaporation experiment

3.1 Methods

After initial tests targeting the functional capabilities of the sampler (timing, rotation angles, etc.), we studied the evaporation reduction efficiency of the device.

To this end, five HDPE bottles (500 mL; Labsolute, Renningen, Germany) were partially filled with water of known isotopic composition (δ18O=-8.53 ‰, δ2H=-60.7 ‰ related to Vienna Standard Mean Ocean Water, V-SMOW) and connected with water and gas tubing to the distribution unit (Table 1). The latter was not coupled to the microcontroller unit; i.e., the rotor did not move during the test.

Table 1Overview of collected samples and associated water losses and isotopic shifts. Water losses Δm are given as absolute mass losses (g) and as percentages of the original water amount (% orig). Isotopic signatures (δ18O and δ2H) and isotopic shifts away from the original composition (Δδ18O and Δδ2H) are given in ‰ V-SMOW. Note that the analytical precisions for δ18O and δ2H account for ±0.15 ‰ and ±0.6 ‰ (±1σ), respectively.

n/a: not applicable. * Δδ within analytical precision (|±1σ|=2σ).

Download Print Version | Download XLSX

Bottle 1 (filled with 100 mL) was connected to the open ports; i.e., its water and air tubes led to the ports beneath the rain inlet and the air outlet, respectively (Fig. 2). A 2 m long rain tube (without funnel) was connected to the rain inlet, and a 15 m long pressure equilibration tube was connected to the air outlet. With this bottle, we wanted to assess how long a sampling bottle can be left “exposed” to the atmosphere (exposed via the rain tube and the pressure equilibration tube).

https://www.hydrol-earth-syst-sci.net/23/2637/2019/hess-23-2637-2019-f02

Figure 2Photograph of the evaporation experiment setup. Note that the bottles were pre-filled with water of known isotopic composition; i.e., no water flowed through the tubing, and the distribution unit did not rotate during the test.

Download

Bottles 2 to 5 contained different water amounts (100, 200, 300, and 400 mL), and their water and air tubes led to stator ports that were blocked by the rotor disc. Here, the goal was to determine how long a sample can remain in the collector without undergoing critical evaporative mass loss.

This setup was placed for 26 weeks (see Hartmann et al., 2018) in a laboratory drying oven (T 6120 by Heraeus, Hanau, Germany) with which a diurnal temperature regime was simulated (Michelsen et al., 2018). A socket timer triggered a daily 12 h heating period (31 C). After this phase, the oven was allowed to cool down to room temperature (approx. 21 C). To accelerate this cooling process, the oven was opened daily for 3 h. Temperature and relative humidity in the oven were logged in 10 min intervals (DK320 HumiLog ruggedPlus by Driesen + Kern GmbH, Bad Bramstedt, Germany).

For comparison, two additional bottles were placed into the oven – bottles 6 and 7 (both filled with 100 mL). Bottle 6 was closed, thus representing a best-case scenario. With this approach, potential diffusion through the bottle material was tested. Bottle 7, by contrast, was “open”; i.e., it featured two holes (diameter 6 mm) in its cap, to represent the worst case (no evaporation suppression).

Moreover, three identical LDPE tubing loops (häberle LABORTECHNIK, Lonsee-Ettlenschieß, Germany) were included in the oven experiment. They were partially filled with 0.5 mL of water and the tubing ends were connected to each other with a connector. All three loops had a circumference of 25 cm, which corresponds to the lengths of the tubing between the bottles and the distribution unit. This allowed for an estimate of diffusive water fluxes through the tubing material.

To determine evaporative mass losses during the experiment, the bottles and the tubing loops were weighed repeatedly (Voyager Pro VP2102C by Ohaus, Pine Brook, USA). Bottles 1 through 5 were disconnected from the distribution unit for this purpose. Samples for isotope analyses (1.6 mL in 2 mL vials) were gathered after 6, 16, and 26 weeks. In the case of bottles 1 to 5, water was withdrawn through the water tube with a syringe. Bottle 6 was sampled through a gastight sampling port in its cap (Mininert by VICI Precision Sampling, Baton Rouge, LA, USA) with a syringe. Samples from bottle 7 were taken with a Pasteur pipette through one of the holes in the cap. To account for these artificial mass losses, the bottles were weighed before and after sampling.

At the end of the test, the obtained water samples were analyzed for their isotopic composition using laser cavity ring-down spectroscopy (L2130-i by Picarro, Santa Clara, CA, USA). The results were expressed in per mil (‰) using the conventional delta notation relative to V-SMOW. The external precisions (±1σ), determined by repeated analyses of a control sample, were ±0.15 ‰ and ±0.6 ‰, for δ18O and δ2H, respectively.

3.2 Results

3.2.1 Evaporative mass losses

The applied diurnal temperature regime (21–31 C; see Fig. S3) mostly resulted in small but measurable evaporative mass losses that increased over time (Table 1).

After 26 weeks, absolute mass losses ranged between 0.8 and 2.4 g for the bottles connected to the distribution unit (bottles 1 through 5). Bottle 6 (closed), representing the best case, showed a lower mass loss of 0.24 g, and in the case of bottle 7 (open; worst case), a loss of 24.42 g was encountered. The three tubing loops (each 25 cm long) lost about 0.14 g over the same time period (see Table S2).

These data suggest that the diffusive loss through the tubing material of the connected bottles (two tubes per bottle, hence 0.28 g; see tubing loop data) is similar to the flux through the bottle material of bottle 6 (0.24 g). As all connected bottles exhibited greater absolute mass losses, additional leakages, e.g., at the cable grommets in the bottle caps or at the distribution unit, seem likely. In this context, pressure fluctuations, induced by the diurnal temperature regime, probably play a role. It is also noteworthy that the blocked bottles, bottles 4 and 5, containing 300 and 400 mL of water, showed the greatest losses (>2 g). This observation could point towards an influence of the bottle surface area in contact with liquid water on the diffusive water flux through the plastic. Nevertheless, the overall absolute losses are still rather small, particularly when compared to the worst-case scenario, an unprotected bottle (bottle 7).

This becomes clearer when the data are put in perspective with the original water amounts. Fractional losses of bottles 1 through 5 range from 0.40 % to 0.94 % orig; i.e., even after half a year the maximum loss was below 1 %. As expected, these values are somewhat higher than in case of the closed bottle (bottle 6; 0.24 % orig) but far below the loss recorded for the open bottle, bottle 7, which lost nearly a quarter of its water (24.42 % orig).

3.2.2 Isotopic shifts

The bulk of the obtained δ values scatter around the original isotopic signature (δ18O=-8.53 ‰, δ2H=-60.7 ‰; see Table 1). The calculated isotopic shifts, Δδ18O and Δδ2H, mostly range between −0.07 ‰ and 0.30 ‰ and between −0.2 ‰ and 0.7 ‰, respectively (bottles 1 through 5).

These shifts are rather small. Keeping in mind that the reported external 1σ precisions, ±0.15 ‰ (δ18O) and ±0.6 ‰ (δ2H), apply to the original water and the water after oven exposure, the encountered shifts practically all lie within the analytical error. This also holds true for the shifts of the closed bottle, bottle 6.

The open bottle, bottle 7, by contrast, showed substantial shifts. After 26 weeks, the δ18O and δ2H values had increased by about 9.23 ‰ and 25.5 ‰, respectively.

4 Discussion

Following a “keep it simple” approach, we have designed an elementary and robust sampler, deliberately avoiding complexity such as pumps, solenoid valves, or remote controls (cf. Coplen et al., 2015; Hartmann et al., 2018; Nelke et al., 2018). Avoiding such complexities greatly reduces the failure risk in the field. We also avoided components that permanently consume power (e.g., normally closed solenoid valves during the “open” phase) and complex tubing systems that would possibly require parts of the sampling water to be used for tubing flushes. Hence, our collector is relatively easy to reproduce, although some technical skills are required.

Compared to other devices (Coplen et al., 2008; Hartmann et al., 2018), it has fewer but bigger bottles. The latter aspect implies that analyses do not have to be restricted to δ18O and δ2H, but other parameters such as 3H and major ions (incl. bicarbonate by titration), etc., could in principle be studied as well.

Most importantly, our device gathers integral samples over freely selectable collection periods (minutes to weeks) and effectively reduces post-sampling evaporation. In combination with the low power consumption, these features render it a potential candidate for autonomous rain sampling at remote, unmanned sites. Traditionally, precipitation samples for isotope analyses are mostly collected on a monthly integral basis (e.g., in the Global Network of Isotopes in Precipitation, GNIP; IAEA/WMO, 2019). Yet, due to the advent of laser-based isotope analyzers and the associated decreasing costs for analyses (Berman et al., 2009; Herbstritt et al., 2012; Wassenaar et al., 2018), shorter collection periods (e.g., weekly integral samples) have become more popular (e.g., Otte et al., 2017). An important advantage of sampling schemes with a higher temporal resolution is the gain in flexibility. They allow, for example, a correlation between isotope signature and meteorological variables (Akers et al., 2017; Hughes and Crawford, 2013; Rao et al., 2008), but data can still be aggregated to precipitation-weighted monthly values for seamless comparison with other monthly data sets (e.g., GNIP).

Due to the relatively low costs (parts < EUR 600), several automatic collectors can be deployed simultaneously to address spatial variability. A number of researchers reported pronounced spatial variability of the isotopic composition of rain in small catchments (Fischer et al., 2017; Kato et al., 2013) or at the city scale (Chen et al., 2017). In these studies, δ18O values partly varied by several ‰ within short distances (a few hundred meters to a few kilometers). In the case of throughfall, δ18O values can even differ significantly on the plot scale, i.e., within meters, and differences of more than 1 ‰ (Allen et al., 2014, 2015; Kato et al., 2013) or even several ‰ (Hsueh et al., 2016) have been observed. Further, it has been shown that ignoring the spatial variability impairs isotope-based hydrograph separations (Cayuela et al., 2019; Fischer et al., 2017). These findings highlight the need for collector arrays or networks (see Lutz et al., 2018; Scholl et al., 1995, 1996). Using expensive, commercial equipment in such networks might be prohibitively expensive for many researchers. Rodgers et al. (2005), for instance, report resource constraints preventing the installation of a second sampler in their catchment to capture the local altitude effect. In such cases, an affordable custom-made alternative is extremely useful. This also applies to study areas in which vandalism or theft of monitoring equipment might be an issue (see Kongo et al., 2010; Otte et al., 2017; Pramana and Ertsen, 2016). In such settings, researchers on a limited budget would possibly hesitate to leave expensive equipment unattended at a remote site for longer time periods, and a low-cost alternative would be appreciated.

Although this work focused on the isotopes 18O and 2H, gathered samples could also be analyzed for other isotopes or their hydrochemistry. Since the funnel is permanently exposed, the sampler would act as a bulk sampler (collecting wet and dry deposition). Hydrochemical analyses could, for instance, include chloride for recharge estimations via the chloride mass balance method (e.g., Eriksson and Khunakasem, 1969; Guan et al., 2010).

5 Potential modifications

While the presented sampler suits our purposes, we acknowledge that other researchers might have different expectations towards such a collector. Thus, we explicitly encourage others to modify our design and tailor it to their specific needs.

Potential modifications might address the number of bottles and their size. In addition, the tubing can be changed, but we suggest the use of opaque tubing for the exposed section from the funnel to the sampler itself to reduce the risk of algae growth, which has occasionally caused problems in rain collectors (Scholl et al., 1995). Although it might be tempting to replace the push-in ports by ordinary barbed hose fittings, we suggest not to do so for the following reasons. (1) Given the potential risk of disconnection of tubing, the used ports are deemed a safe option. (2) Their handling is easier, particularly when dealing with tens of connections in a confined space. (3) They do not introduce an additional constriction.

Although we used a Texas Instruments microcontroller, we can imagine that the functionality of the sampler could also be achieved with parts from the popular Arduino ecosystem, though probably at the cost of greater power consumption.

Moreover, one could transform the timer-actuated into a volume-controlled rain sampler, for instance by means of a microcontroller-based tipping bucket system. The latter would calculate the filling status of a sampling bottle based on the recorded number of tips and automatically direct the water into the next bottle when the first one is full (see Muller et al., 2015).

Also, a transformation into a surface water sampler is feasible. To this end, one could combine the current device with a timer-triggered peristaltic or submersible pump. The microcontroller provides several unused input/output and communication pins and further unutilized resources to allow for such customizations.

6 Summary and conclusions

Our microcontroller-based automatic rain sampler enables timer-actuated integral rain sampling. The simple, low-cost device is robust and effectively minimizes post-sampling evaporation from the collection bottles and the associated isotope fractionation. The excellent performance of the device during an extensive evaporation experiment in a laboratory oven (26 weeks; 21–31 C) suggests that even multi-week field deployments in warm climates are feasible. In the spirit of open science, we share all relevant details on our sampler and encourage others to adapt it to their specific needs.

Data availability

All details needed to copy our sampler are freely available (Supplement and https://www.ufz.de/index.php?en=44048, last access: 15 June 2019, section Documentation). Data on the evaporation experiment, enabling a performance evaluation, are given in the main article and the Supplement.

Supplement

The supplement related to this article is available online at: https://doi.org/10.5194/hess-23-2637-2019-supplement.

Author contributions

NM, GL, and TM conceptualized the sampler. GL was responsible for the detailed sampler design and programmed the microcontroller. GL, JF, ABABS, and TM conducted initial tests. NM carried out the evaporation experiment. SMW performed isotope and data analyses. NM wrote the manuscript with contributions from all co-authors.

Competing interests

The authors declare that they have no conflict of interest.

Acknowledgements

This work has been supported by the project “Submarine Groundwater Discharge: Adaption of an Autonomous Aquatic Vehicle for Robotic Measurements, Sampling and Monitoring”, funded by The Research Council of Oman (TRC Research Contract No. TRC/RCP/15/001). We would also like to thank the UFZ workshop and Hendrik Zöphel (UFZ) for assistance in construction and Inga Schreiter, Sahand Farhang Darehshouri, and Claudia Cosma (TU Darmstadt) for help during the evaporation experiment. Moreover, we thank Martijn Westhoff for his efficient editorial handling and Rolf Hut and Manfred Gröning for their constructive reviews. We acknowledge support by the German Research Foundation and the Open Access Publishing Fund of Technische Universität Darmstadt.

Financial support

This research has been supported by The Research Council of Oman (grant no. TRC/RCP/15/001).

Review statement

This paper was edited by Martijn Westhoff and reviewed by Manfred Groening and Rolf Hut.

References

Akers, P. D., Welker, J. M., and Brook, G. A.: Reassessing the role of temperature in precipitation oxygen isotopes across the eastern and central United States through weekly precipitation-day data, Water Resour. Res., 53, 7644–7661, https://doi.org/10.1002/2017WR020569, 2017. 

Allen, S. T., Brooks, J. R., Keim, R. F., Bond, B. J., and McDonnell, J. J.: The role of pre-event canopy storage in throughfall and stemflow by using isotopic tracers, Ecohydrol., 7, 858–868, https://doi.org/10.1002/eco.1408, 2014. 

Allen, S. T., Keim, R. F., and McDonnell, J. J.: Spatial patterns of throughfall isotopic composition at the event and seasonal timescales, J. Hydrol., 522, 58–66, https://doi.org/10.1016/j.jhydrol.2014.12.029, 2015. 

Ankor, M. J., Tyler, J. J., and Hughes, C. E.: Development of an autonomous, monthly and daily, rainfall sampler for isotope research, J. Hydrol., 575, 31–41, https://doi.org/10.1016/j.jhydrol.2019.04.074, 2019. 

Berman, E. S. F., Gupta, M., Gabrielli, C., Garland, T., and McDonnell, J. J.: High-frequency field-deployable isotope analyzer for hydrological applications, Water Resour. Res., 45, W10201, https://doi.org/10.1029/2009WR008265, 2009. 

Birkel, C., Tetzlaff, D., Dunn, S. M., and Soulsby, C.: Using lumped conceptual rainfall–runoff models to simulate daily isotope variability with fractionation in a nested mesoscale catchment, Adv. Water Resour., 34, 383–394, https://doi.org/10.1016/j.advwatres.2010.12.006, 2011. 

Bowen, G. J. and Revenaugh, J.: Interpolating the isotopic composition of modern meteoric precipitation, Water Resour. Res., 39, 1299, https://doi.org/10.1029/2003WR002086, 2003. 

Buda, A. R. and DeWalle, D. R.: Dynamics of stream nitrate sources and flow pathways during stormflows on urban, forest and agricultural watersheds in central Pennsylvania, USA, Hydrol. Process., 23, 3292–3305, https://doi.org/10.1002/hyp.7423, 2009. 

Camacho Suarez, V. V., Saraiva Okello, A. M. L., Wenninger, J. W., and Uhlenbrook, S.: Understanding runoff processes in a semi-arid environment through isotope and hydrochemical hydrograph separations, Hydrol. Earth Syst. Sc., 19, 4183–4199, https://doi.org/10.5194/hess-19-4183-2015, 2015. 

Cayuela, C., Latron, J., Geris, J., and Llorens, P.: Spatio-temporal variability of the isotopic input signal in a partly forested catchment: Implications for hydrograph separation, Hydrol. Process., 33, 36–46, https://doi.org/10.1002/hyp.13309, 2019. 

Chen, F., Zhang, M., Wang, S., Qiu, X., and Du, M.: Environmental controls on stable isotopes of precipitation in Lanzhou, China: An enhanced network at city scale, Sci. Total Environ., 609, 1013–1022, https://doi.org/10.1016/j.scitotenv.2017.07.216, 2017. 

Conroy, J. L., Noone, D., Cobb, K. M., Moerman, J. W., and Konecky, B. L.: Paired stable isotopologues in precipitation and vapor: A case study of the amount effect within western tropical Pacific storms, J. Geophys. Res.-Atmos., 121, 3290–3303, https://doi.org/10.1002/2015JD023844, 2016. 

Coplen, T. B.: Sequential, time-integrated collector of precipitation, ground water, and surface water for analysis of isotopes, Patent US7687028 (B1), 2010. 

Coplen, T. B., Neiman, P. J., White, A. B., Landwehr, J. M., Ralph, M., and Dettinger, M. D.: Extreme changes in stable hydrogen isotopes and precipitation characteristics in a landfalling Pacific storm, Geophys. Res. Lett., 35, L21808, https://doi.org/10.1029/2008GL035481, 2008. 

Coplen, T. B., Neiman, P. J., White, A. B., and Ralph, F. M.: Categorisation of northern California rainfall for periods with and without a radar brightband using stable isotopes and a novel automated precipitation collector, Tellus B, 67, 28574, https://doi.org/10.3402/tellusb.v67.28574, 2015. 

Eriksson, E. and Khunakasem, V.: Chloride concentration in groundwater, recharge rate and rate of deposition of chloride in the Isreal coastal plan, J. Hydrol., 7, 178–197, https://doi.org/10.1016/0022-1694(69)90055-9, 1969. 

Fergusson, S. P.: Improved Gages for Precipitation, Mon. Weather Rev., 49, 379–386, https://doi.org/10.1175/1520-0493(1921)49<379:IGFP>2.0.CO;2, 1921. 

Fischer, B. M. C., Stähli, M., and Seibert, J.: Pre-event water contributions to runoff events of different magnitude in pre-alpine headwaters, Hydrol. Res., 48, 28–47, https://doi.org/10.2166/nh.2016.176, 2016. 

Fischer, B. M. C., van Meerveld, H. J., and Seibert, J.: Spatial variability in the isotopic composition of rainfall in a small headwater catchment and its effect on hydrograph separation, J. Hydrol., 547, 755–769, https://doi.org/10.1016/j.jhydrol.2017.01.045, 2017. 

Gröning, M.: Interactive comment on “Technical note: A microcontroller-based automatic rain sampler for stable isotope studies” by N. Michelsen et al., https://doi.org/10.5194/hess-2019-93-RC2, 2019. 

Gröning, M., Lutz, H. O., Roller-Lutz, Z., Kralik, M., Gourcy, L., and Pöltenstein, L.: A simple rain collector preventing water re-evaporation dedicated for δ18O and δ2H analysis of cumulative precipitation samples, J. Hydrol., 448–449, 195–200, https://doi.org/10.1016/j.jhydrol.2012.04.041, 2012. 

Guan, H., Love, A. J., Simmons, C. T., Makhnin, O., and Kayaalp, A. S.: Factors influencing chloride deposition in a coastal hilly area and application to chloride deposition mapping, Hydrol. Earth Syst. Sci., 14, 801–813, https://doi.org/10.5194/hess-14-801-2010, 2010. 

Hartmann, A., Luetscher, M., Wachter, R., Holz, P., Eiche, E., and Neumann, T.: Technical note: GUARD – an automated fluid sampler preventing sample alteration by contamination, evaporation and gas exchange, suitable for remote areas and harsh conditions, Hydrol. Earth Syst. Sci., 22, 4281–4293, https://doi.org/10.5194/hess-22-4281-2018, 2018. 

Herbstritt, B., Gralher, B., and Weiler, M.: Continuous in situ measurements of stable isotopes in liquid water, Water Resour. Res., 48, W03601, https://doi.org/10.1029/2011WR011369, 2012. 

Herbstritt, B., Gralher, B., and Weiler, M.: Real-time observations of stable isotope dynamics during rainfall and throughfall events, Hydrol. Earth Syst. Sci. Discuss., https://doi.org/10.5194/hess-2018-301, in review, 2018a. 

Herbstritt, B., Seeger, S., Rinderer, M., and Weiler, M.: Low cost water vapour sampling for mobile in-situ measurements of stable water isotopes, EGU General Assembly 2018, Vienna, Austria, EGU2018-9188-1, 2018b. 

Hervé-Fernández, P., Oyarzún, C., Brumbt, C., Huygens, D., Bodé, S., Verhoest, N. E. C., and Boeckx, P.: Assessing the `two water worlds' hypothesis and water sources for native and exotic evergreen species in south-central Chile, Hydrol. Process., 30, 4227–4241, https://doi.org/10.1002/hyp.10984, 2016. 

Hsueh, Y. H., Allen, S. T., and Keim, R. F.: Fine-scale spatial variability of throughfall amount and isotopic composition under a hardwood forest canopy, Hydrol. Process., 30, 1796–1803, https://doi.org/10.1002/hyp.10772, 2016. 

Hughes, C. E. and Crawford, J.: Spatial and temporal variation in precipitation isotopes in the Sydney Basin, Australia, J. Hydrol., 489, 42–55, https://doi.org/10.1016/j.jhydrol.2013.02.036, 2013. 

IAEA: IAEA/GNIP precipitation sampling guide, 2014. 

IAEA/WMO: Global Network of Isotopes in Precipitation., The GNIP Database, available at: https://nucleus.iaea.org/wiser, last access: 12 January 2019. 

Kato, H., Onda, Y., Nanko, K., Gomi, T., Yamanaka, T., and Kawaguchi, S.: Effect of canopy interception on spatial variability and isotopic composition of throughfall in Japanese cypress plantations, J. Hydrol., 504, 1–11, https://doi.org/10.1016/j.jhydrol.2013.09.028, 2013. 

Kennedy, V. C., Zellweger, G. W., and Avanzino, R. J.: Variation of Rain Chemistry During Storms at Two Sites in Northern California, Water Resour. Res., 15, 687–702, https://doi.org/10.1029/WR015i003p00687, 1979. 

Kongo, V. M., Kosgei, J. R., Jewitt, G. P. W., and Lorentz, S. A.: Establishment of a catchment monitoring network through a participatory approach in a rural community in South Africa, Hydrol. Earth Syst. Sci., 14, 2507–2525, https://doi.org/10.5194/hess-14-2507-2010, 2010. 

Krupa, S. V.: Sampling and physico-chemical analysis of precipitation: a review, Environ. Pollut., 120, 565–594, https://doi.org/10.1016/S0269-7491(02)00165-3, 2002. 

Laquer, F. C.: Sequential Precipitation Samplers: A Literature Review, Atmos. Environ., 24A, 2289–2297, https://doi.org/10.1016/0960-1686(90)90322-E, 1990. 

Lutz, S. R., Krieg, R., Müller, C., Zink, M., Knöller, K., Samaniego, L., and Merz, R.: Spatial Patterns of Water Age: Using Young Water Fractions to Improve the Characterization of Transit Times in Contrasting Catchments, Water Resour. Res., 54, 4767–4784, https://doi.org/10.1029/2017WR022216, 2018. 

Michelsen, N., Reshid, M., Siebert, C., Schulz, S., Knöller, K., Weise, S. M., Rausch, R., Al-Saud, M., and Schüth, C.: Isotopic and chemical composition of precipitation in Riyadh, Saudi Arabia, Chem. Geol., 413, 51–62, https://doi.org/10.1016/j.chemgeo.2015.08.001, 2015. 

Michelsen, N., van Geldern, R., Roßmann, Y., Bauer, I., Schulz, S., Barth, J. A. C., and Schüth, C.: Comparison of cumulative precipitation collectors used in isotope hydrology, Chem. Geol, 488, 171–179, https://doi.org/10.1016/j.chemgeo.2018.04.032, 2018. 

Muller, C. L., Baker, A., Fairchild, I. J., Kidd, C., and Boomer, I.: Intra-Event Trends in Stable Isotopes: Exploring Midlatitude Precipitation Using a Vertically Pointing Micro Rain Radar, J. Hydrometeorol., 16, 194–213, https://doi.org/10.1175/JHM-D-14-0038.1, 2015. 

Munksgaard, N. C., Wurster, C. M., and Bird, M. I.: Continuous analysis of δ18O and δD values of water by diffusion sampling cavity ring-down spectrometry: a novel sampling device for unattended field monitoring of precipitation, ground and surface waters, Rapid Commun. Mass Sp., 25, 3706–3712, https://doi.org/10.1002/rcm.5282, 2011. 

Munksgaard, N. C., Wurster, C. M., Bass, A., and Bird, M. I.: Extreme short-term stable isotope variability revealed by continuous rainwater analysis, Hydrol. Process., 26, 3630–3634, https://doi.org/10.1002/hyp.9505, 2012. 

Muñoz-Villers, L. E. and McDonnell, J. J.: Land use change effects on runoff generation in a humid tropical montane cloud forest region, Hydrol. Earth Syst. Sci., 17, 3543–3560, https://doi.org/10.5194/hess-17-3543-2013, 2013. 

Nelke, M., Coltman, C., Lien, H., Walter, C., Udell, C., and Selker, J.: OPEnSLab-OSU/OPEnSampler: OPEnSampler v1.0, https://doi.org/10.5281/zenodo.2483259 (see also http://www.open-sensing.org/opensampler/, last access: 10 February 2019), 2018. 

Neuhaus, L.: Verfahren mit einem elektronischen Probenfluidnehmer zum Erzeugen und Bereitstellen von in umgebungsluftdichten Behältern abgefüllten Fluidproben für die Analyse und/oder Grenzwert-Überwachung, Patent DE102015007049 (A1), 2016. 

Neuhaus, L.: Automatisches Wasserprobenentnahmesystem mit Verdunstungs- und Entgasungsschutz, available at: https://www.liquidsampler.de/, last access: 12 January 2019. 

Otte, I., Detsch, F., Gütlein, A., Scholl, M., Kiese, R., Appelhans, T., and Nauss, T.: Seasonality of stable isotope composition of atmospheric water input at the southern slopes of Mt. Kilimanjaro, Tanzania, Hydrol. Process., 31, 3932–3947, https://doi.org/10.1002/hyp.11311, 2017. 

Pangle, L. A., Klaus, J., Berman, E. S. F., Gupta, M., and McDonnell, J. J.: A new multisource and high-frequency approach to measuring δ2H and δ18O in hydrological field studies, Water Resour. Res., 49, 7797–7803, https://doi.org/10.1002/2013WR013743, 2013. 

Pramana, K. E. R. and Ertsen, M. W.: Towards systematic planning of small-scale hydrological intervention-based research, Hydrol. Earth Syst. Sci., 20, 4093–4115, https://doi.org/10.5194/hess-20-4093-2016, 2016. 

Qu, S., Wang, Y., Zhou, M., Liu, H., Shi, P., Yu, Z., and Xiang, L.: Temporal 18O and deuterium variations in hydrologic components of a small watershed during a typhoon event, Isot. Environ. Healt. S., 53, 172–183, https://doi.org/10.1080/10256016.2016.1205590, 2017. 

Rao, T. N., Radhakrishna, B., Srivastava, R., Satyanarayana, T. M., Rao, D. N., and Ramesh, R.: Inferring microphysical processes occurring in mesoscale convective systems from radar measurements and isotopic analysis, Geophys. Res. Lett., 35, L09813, https://doi.org/10.1029/2008GL033495, 2008. 

Robertson, J. K., Dolzine, T. W., and Graham, R. C.: Chemistry of Precipitation from Sequentially Sampled Storms, U.S. Environmental Protection Agency, EPA-600/4-80-004, 1980. 

Rodgers, P., Soulsby, C., Waldron, S., and Tetzlaff, D.: Using stable isotope tracers to assess hydrological flow paths, residence times and landscape influences in a nested mesoscale catchment, Hydrol. Earth Syst. Sci., 9, 139–155, https://doi.org/10.5194/hess-9-139-2005, 2005. 

Saffarpour, S., Western, A. W., Adams, R., and McDonnell, J. J.: Multiple runoff processes and multiple thresholds control agricultural runoff generation, Hydrol. Earth Syst. Sci., 20, 4525–4545, https://doi.org/10.5194/hess-20-4525-2016, 2016. 

Scholl, M. A., Ingebritsen, S. E., Janik, C. J., and Kauahikaua, J. P.: An Isotope Hydrology Study of the Kilauea Volcano Area, Hawaii, U.S. Geological Survey Water-Resources Investigations Report 95-4213, 1995. 

Scholl, M. A., Ingebritsen, S. E., Janik, C. J., and Kauahikaua, J. P.: Use of precipitation and groundwater isotopes to interpret regional hydrology on a tropical volcanic island: Kilauea volcano area, Hawaii, Water Resour. Res., 32, 3525–3537, https://doi.org/10.1029/95WR02837, 1996. 

Siebert, C.: Automatischer Probennehmer mit atmosphärisch entkoppelter Speicherung wässriger Proben im Passiv- und Aktiv-Modus, Patent DE102014205633 (B4), 2015. 

Spangenberg, J. E.: Caution on the storage of waters and aqueous solutions in plastic containers for hydrogen and oxygen stable isotope analysis, Rapid Commun. Mass Sp., 26, 2627–2636, https://doi.org/10.1002/rcm.6386, 2012.  

Sprenger, M., Tetzlaff, D., and Soulsby, C.: Soil water stable isotopes reveal evaporation dynamics at the soil–plant–atmosphere interface of the critical zone, Hydrol. Earth Syst. Sci., 21, 3839–3858, https://doi.org/10.5194/hess-21-3839-2017, 2017. 

Tauro, F., Selker, J., van de Giesen, N., Abrate, T., Uijlenhoet, R., Porfiri, M., Manfreda, S., Caylor, K., Moramarco, T., Benveniste, J., Ciraolo, G., Estes, L., Domeneghetti, A., Perks, M. T., Corbari, C., Rabiei, E., Ravazzani, G., Bogena, H., Harfouche, A., Brocca, L., Maltese, A., Wickert, A., Tarpanelli, A., Good, S., Lopez Alcala, J. M., Petroselli, A., Cudennec, C., Blume, T., Hut, R., and Grimaldi, S.: Measurements and Observations in the XXI century (MOXXI): innovation and multi-disciplinarity to sense the hydrological cycle, Hydrolog. Sci. J., 63, 169–196, https://doi.org/10.1080/02626667.2017.1420191, 2018. 

Terzer-Wassmuth, S., Araguas-Araguas, L. J., Wassenaar, L. I., and Aggarwal, P. K.: On the short-term (5-30 min.) precipitation isotope (δ18O, δ2H) variability in a temperate climate, EGU General Assembly 2018, Vienna, Austria, EGU2018-18997, 2018. 

Tunaley, C., Tetzlaff, D., Birkel, C., and Soulsby, C.: Using high-resolution isotope data and alternative calibration strategies for a tracer-aided runoff model in a nested catchment, Hydrol. Process., 31, 3962–3978, https://doi.org/10.1002/hyp.11313, 2017. 

van Huijgevoort, M. H. J., Tetzlaff, D., Sutanudjaja, E. H., and Soulsby, C.: Using high resolution tracer data to constrain water storage, flux and age estimates in a spatially distributed rainfall-runoff model, Hydrol. Process., 30, 4761–4778, https://doi.org/10.1002/hyp.10902, 2016. 

Vermette, S. J. and Drake, J. J.: Simplified Wet-Only and Sequential Fraction Rain Collector, Atmos. Environ., 21, 715–716, https://doi.org/10.1016/0004-6981(87)90053-9, 1987. 

von Freyberg, J., Studer, B., and Kirchner, J. W.: A lab in the field: high-frequency analysis of water quality and stable isotopes in stream water and precipitation, Hydrol. Earth Syst. Sci., 21, 1721–1739, https://doi.org/10.5194/hess-21-1721-2017, 2017. 

Wassenaar, L. I., Terzer-Wassmuth, S., Douence, C., Araguas-Araguas, L., Aggarwal, P. K., and Coplen, T. B.: Seeking excellence: an evaluation of 235 international laboratories conducting water isotope analyses by isotope-ratio and laser-absorption spectrometer, Rapid Commun. Mass Sp., 32, 393–406, https://doi.org/10.1002/rcm.8052, 2018. 

Williams, M. R., Lartey, J. L., and Sanders, L. L.: Isotopic (δ18O and δ2H) Integrity of Water Samples Collected and Stored by Automatic Samplers, Agric. Environ. Lett., 3, 180009, https://doi.org/10.2134/ael2018.02.0009, 2018. 

Windhorst, D., Kraft, P., Holly, H., Sahraei, A., and Breuer, L.: A mobile and self-sufficient lab for high frequency measurements of stable water isotopes and chemistry of multiple water sources, EGU General Assembly 2017, Vienna, Austria, EGU2017-4879, 2017. 

Zannoni, D., Steen-Larsen, H. C., Rampazzo, G., Dreossi, G., Stenni, B., and Bergamasco, A.: The atmospheric water cycle of a coastal lagoon: An isotope study of the interactions between water vapor, precipitation and surface waters, J. Hydrol., 572, 630–644, https://doi.org/10.1016/j.jhydrol.2019.03.033, 2019. 

Download
Short summary
Most commercial automatic rain samplers are costly and do not prevent evaporation from the collection bottles. Hence, we have developed a microcontroller-based collector enabling timer-actuated integral rain sampling. The simple, low-cost device is robust and effectively minimizes post-sampling evaporation. The excellent performance of the collector during an evaporation experiment in a lab oven suggests that even multi-week field deployments in warm climates are feasible.