Articles | Volume 22, issue 7
https://doi.org/10.5194/hess-22-3619-2018
https://doi.org/10.5194/hess-22-3619-2018
Research article
 | Highlight paper
 | 
06 Jul 2018
Research article | Highlight paper |  | 06 Jul 2018

Inter-laboratory comparison of cryogenic water extraction systems for stable isotope analysis of soil water

Natalie Orlowski, Lutz Breuer, Nicolas Angeli, Pascal Boeckx, Christophe Brumbt, Craig S. Cook, Maren Dubbert, Jens Dyckmans, Barbora Gallagher, Benjamin Gralher, Barbara Herbstritt, Pedro Hervé-Fernández, Christophe Hissler, Paul Koeniger, Arnaud Legout, Chandelle Joan Macdonald, Carlos Oyarzún, Regine Redelstein, Christof Seidler, Rolf Siegwolf, Christine Stumpp, Simon Thomsen, Markus Weiler, Christiane Werner, and Jeffrey J. McDonnell

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
ED: Publish subject to minor revisions (further review by editor) (13 Jun 2018) by Thom Bogaard
AR by Natalie Orlowski on behalf of the Authors (14 Jun 2018)  Author's response    Manuscript
ED: Publish as is (21 Jun 2018) by Thom Bogaard
Download
Short summary
To extract water from soils for isotopic analysis, cryogenic water extraction is the most widely used removal technique. This work presents results from a worldwide laboratory intercomparison test of cryogenic extraction systems. Our results showed large differences in retrieved isotopic signatures among participating laboratories linked to interactions between soil type and properties, system setup, extraction efficiency, extraction system leaks, and each lab’s internal accuracy.